Table 1.
Instrument type |
SSI mean [min– max] |
BNR mean [min–max] |
Qualified instruments |
|
---|---|---|---|---|
SSC instruments |
Accuri C6 | 0 | ND | 0/2 |
Apogee A50a | 7.7 | 0.03 | 1/1 | |
FACSAria | 5.2 [2.8–7.1] | 0.07 [0–0.22] | 4/5 | |
FACSCalibur | 4.6 [3–6.5] | 0.58 [0–2.59] | 5/6 | |
FACSCanto | 4.5 [3.4–7] | 0.07 [0.01–0.2] | 11/11 | |
FACSVerse | 7.1 [7–7.3] | 0.12 [0.06–0.19] | 2/2 | |
Influxa | 2.4 | 0.43 | 0/1 | |
LSR Fortessa | 8 [7.2–9.1] | 0.02 [0.02–0.05] | 3/3 | |
LSR II | 7.9 [5.7–10] | 0.02 [0–0.03] | 4/4 | |
Stratedigma | 4.3 | 0.02 | 1/1 | |
FSC instruments |
Apogee A50a | 11.3 | 0 | 1/1 |
Epics XL | 0 | ND | 0/1 | |
FC500 | 0 | ND | 0/2 | |
Guava | 0 | ND | 0/2 | |
Influxa | 11.3 | 0.01 | 1/1 | |
Navios/Gallios | 5.6 [2.8–7.6] | 1.53 [0–6.2] (0.03)b |
7/12 (12/12)b | |
Stratedigma | 4 | 1 | 1/1 | |
Total | 39/52 (44/52)b |
Instruments tested both in SSC and FSC;
Results after flow cell wash;
Sensitivity index (SSI) = (Median bead A− Median bead B)/ (SD bead A+ SD bead B) where bead A = 0.2µm and bead B = 0.16µm for SSC FCMrs, and bead A = 0.5 µm and bead B = 0.3 µm for FSC FCMrs. SSI > 3 was required to be compatible with the standardization strategy. Background noise ratio (BNR) is the ratio between the number of events per second measured in the protocol settings and the instrument specific maximal number of events per second keeping abort rate at a low level (FACSCanto I/II = 4000, FACSCalibur = 2000, FACSAria I/II = 4000, LSRII (+/− Fortessa) = 4000, FACSVerse = 4000, Apogee A50 = 2000, Stratedigm = 4000, Gallios/Navios = 5000, Influx = 15 000, defined according to both instrument specifications and core lab validation).