Skip to main content
. Author manuscript; available in PMC: 2017 Nov 1.
Published in final edited form as: Anal Chem. 2016 Oct 14;88(21):10687–10692. doi: 10.1021/acs.analchem.6b03191

Table 4.

Intra-day and inter-day accuracy (% bias) and precision (% RSD) of spiked urines at three different concentrations. Samples were analyzed in triplicate. Inter-day data were estimated from spiked urine samples analyzed for five different days over several weeks (n=15)

Intra-Day Inter-Day

Analyte Actual
Concentration
(ng/mL)
Calculated
Concentration
(ng/mL)
Bias
(%)
RSD
(%)
Calculated
Concentration
(ng/mL)
Bias
(%)
RSD
(%)
PPDA 0.50 0.53 6.0 6.0 0.54 8.0 8.30
2.45 2.20 −10.2 4.0 2.78 13.5 6.58
24.5 25.1 2.5 3.7 26.2 6.9 3.58
6TDA 0.48 0.47 −2.1 3.8 0.45 −6.3 7.60
2.45 2.40 −2.0 4.5 2.50 2.0 4.44
24.4 25.2 3.3 1.5 27.4 12.3 2.06
4TDA 0.49 0.44 −10.2 2.0 0.51 4.1 3.42
2.44 2.39 −2.0 2.4 2.52 3.3 4.49
24.4 24.6 0.90 2.7 27.0 10.7 2.59
NDA 0.50 0.47 −6.0 2.3 0.51 2.0 7.14
2.51 2.58 2.8 2.6 2.33 −7.2 2.28
25.1 25.3 0.8 2.7 24.0 −4.4 2.41
MDA 0.16 0.162 1.25 2.7 0.17 6.3 5.09
0.51 0.525 2.94 1.0 0.58 13.7 2.76
5.10 5.06 −0.78 2.5 5.00 −2.0 5.09