Skip to main content
. 2017 Feb 8;17:9. doi: 10.1186/s12886-017-0407-5

Table 1.

Modulation Transfer Function values at 3-mm aperture

MTF Spherical IOL Aspheric IOL P-Value
50 lp/mm Median (Range) 0.7133 (0.6900 to 0.7310) 0.8058 (0.7970 to 0.8130) 0.0002*
100 lp/mm Median (Range) 0.5265 (0.4980 to 0.5430) 0.6345 (0.6230 to 0.6450) 0.0002*

*Statistically significant difference (p < 0.05, Mann-Whitney-U test)