Table 1.

Corresponding Valve-in-Valve matches for different sizes of Stented Bioprosthetic Surgical Aortic Valves

Bioprosthetic Stented Valve True ID (mm) Appropriate TAV Type for ViV Corresponding TAV Annulus Size (mm)
16 – 17.5 Sapien 20 16 – 18
18 – 18.5 Sapien 20 16 – 18
CoreValve 23 18 – 20
Portico 23 19 – 21
19 Sapien 23 18 – 22
CoreValve 23 18 – 20
Portico 23 19 – 21
20.5–21 Sapien 23 18 – 22
CoreValve 23 18 – 20
Portico 23 19 – 21
22 Sapien 26 22 – 25
CoreValve 26 20 – 23
Portico 25 21 – 23
23 Sapien 26 22 – 25
CoreValve 26 20 – 23
Portico 25 21 – 23
24 Sapien 26 22 – 25
CoreValve 26 20 – 23
Portico 27 23 – 25
25–26 Sapien 29 25 – 27.7
CoreValve 29 23 – 27
Portico 27 23 – 25
27 Sapien 29 25 – 27.7
CoreValve 29 23 – 27
Portico 29 25 – 27