. 2017 Feb 10;22(2):028002. doi: 10.1117/1.JBO.22.2.028002

Fig. 2.

Fig. 2

Fluorescence optical property correction phantoms and verification with in-vivo and ex-vivo comparison. (a) Fluorescence SVD amplitude for phantom experiments with varying optical properties and the same BPD concentration (0.25  mg/kg). The best-fit (shown as dashed lines) uses the form a/CF, where CF is Eq. (2) and a=7.645. (Note, CF=1 is normalized for μa=0.69, μs=11  cm1.) (b) BPD concentration (in μM) versus the corrected SVD (SVDcorr). The line of best-fit [BPD]=(0.0301±0.0009)×SVDcorr with R2=0.9986 is used to convert SVDcorr to [BPD] in μM. (c) The measured in-vivo photosensitizer concentration using the multifiber contact probe obtained fluorescence spectra versus ex-vivo measured BPD concentration. The line of best-fit is of the form y=0.9821x with R2=0.9772. The dashed line in (c) represents the line for y=x.