Skip to main content
. 2016 Jan 4;6(1):7. doi: 10.3390/nano6010007

Table 1.

Comparison of the analytical performance of some Bi film-modified electrodes for measurements of Cd(II) and Pb(II).

Electrode Analytical Technique Linear Range (μg·L−1) Detection Limit (μg·L−1) Reference
Cd(II) Pb(II) Cd(II) Pb(II)
Bi/SWNTs/GCE SWASV 0.5–11 0.5–11 0.076 0.18 [33]
Bi/Au-GN-Cys/GCE SWASV 0.50–40 0.50–40 0.10 0.05 [1]
Bi/ABTS-MWCNTs/GCE DPSV 0.5–35 0.2–50 0.2 0.1 [5]
Bi/Nafion/PANI-MES/GCE SWASV 0.1–20 0.1–30 0.04 0.05 [34]
RGO/Bi/GCE SWASV 20–120 20–120 2.8 0.55 [35]
polymer/Bi/GCE SWASV 2–60 2–60 2 2 [36]
Nafion/Bi/GCE SWASV 1–20 1–20 0.1 0.1 [37]
SPAN/MCN/GCE SWASV 5–80 5–80 0.7 0.2 This work

SWNTs: single-walled carbon nanotubes; Au-GN-Cys: gold nanoparticle-graphene-cysteine composite; ABTS: 2,2-azinobis (3-ethylbenzothiazoline-6-sulfonate) diammonium salt; MWCNTs: multi-walled carbon nanotubes; PANI-MES: polyaniline-2-mercaptoethanesulfonate; RGO: reduced graphene oxide.