Table 3.

Labeling results for the clinical validation

Artery Correctly identified existing (TP) Correctly identified non-existing (TN) Not identified (FN) Too short (FN) Sensitivity [%] Specificity [%] Accuracy [%]
Right ICA 10 0 0 0 100 100
Left ICA 10 0 0 0 100 100
BA 10 0 0 0 100 100
Right VA 8 0 0 2 80 80
Left VA 10 0 0 0 100 100
Right PCA 8 0 0 2 80 80
Left PCA 10 0 0 0 100 100
Right MCA 10 0 0 0 100 100
Left MCA 10 0 0 0 100 100
Right ACA 10 0 0 0 100 100
Left ACA 10 0 0 0 100 100
ACAdistal 9 0 1 0 90 90
Right PCoA 0 7 3 0 0 100 70
Left PCoA 0 8 2 0 0 100 80
Right MCAdistal 9 0 1 0 90 90
Left MCAdistal 9 0 1 0 90 90

True positive (TP), false positive (FP), true negative (TN), and false negative (FN) rates, as well as sensitivity and specificity for identifying each artery with AAIM are also presented. Without any TN or FP, it is not possible to calculate specificity

ICA Internal carotid artery, VA Vertebral artery, BA Basilar artery, PCA Posterior cerebral artery, MCA Middle cerebral artery, ACA Anterior cerebral artery, PCoA Posterior communicating artery