Skip to main content
. 2016 Aug 12;7(38):61845–61859. doi: 10.18632/oncotarget.11259

Table 3. Summary of read number, coverage, and mapping results of the two NGS platforms.

Characteristics Illumina
MiSeq
Ion torrent
PGM
Flow cell/chip type Standard 316
Total no. of reads 14M 915k and 1.25M
Mean read length 99.3bp 99.24bp
Average quality of reads (Phred) 33.79 24.24
Average no. of reads/sample 247,749 99,774
%mapped reads 76% 88%
Coverage, mean [min, max] 1,262 [171, 3642] 439 [75, 1326]
% of target area with ≥ 50 fold coverage 99.93 99.68
% of insertions 0.01 1.27
% of deletions 0.04 0.67
% of mismatches 0.29 0.23

All values are average calculations from the 20 analyzed samples.