Table 1.

TI sampling of the ASL signal and its impact on the accuracy of perfusion estimation.

Number of TIs TI values (ms) Mb0∙(1+fss) factor R RPFASL–RPFDCE (ml/min)
16 150, 200, 300, 400, 500, 600, 700, 800, 900, 1000, 1100, 1200, 1300, 1400, 1500, 1600 13.0 ± 3.9 0.92 −9 ±30
14 150, 200, 300, 400, 500, 700, 800, 900, 1000, 1100, 1300, 1400, 1500, 1600 13.2 ± 4.1 0.91 −9 ±31
12 150, 200, 300, 500, 600, 700, 900, 1000, 1100, 1300, 1400, 1500 14.1 ± 5.8 0.85 −9 ± 44
10 150, 300, 500, 600, 800, 1000, 1100, 1300, 1400, 1600 12.9 ± 4.6 0.86 −8 ±45
8 150, 300, 500, 700, 900, 1100, 1300, 1500 13.0 ± 5.6 0.84 −8 ±42
6 150, 400, 700, 1000, 1300, 1600 13.3 ± 3.8 0.86 −12 ± 37
5 150, 500, 800, 1000, 1500 12.2 ± 5.5 0.84 −5 ±51
4 150, 600, 1100, 1600 11.9 ± 6.2 0.80 −12 ± 52
2 400, 1200 19.2 ± 20.3 0.30 16 ± 243