. 2017 Feb 23;12(2):e0172634. doi: 10.1371/journal.pone.0172634

Table 3.

Instrument IOL model Sample size Formula MedAE (D) MAE (D) PE ≤ 0.50 D
Present study Aladdin Acrysof SN60WF 73 Hoffer Q 0.25 0.27 89.04%
Cooke 2016 IOLMaster Acrysof SN60WF 1079 Barrett 0.25 0.31 80.6%
Hoffer 2010 IOLMaster Acrysof SN60WF 50 Haigis N/A 0.46 56%
Olsen IOLMaster AcrySof MA60AC 461 Olsen N/A 0.43 62.5%
Reitblat 2015 IOLMaster Restor SN6AD1 73 SRK/T N/A 0.24 91.78%
Srivannaboon 2013 IOLMaster PY-60AD (Hoya) 163 Hoffer Q N/A 0.32 68%**
Cooke 2016 Lenstar Acrysof SN60WF 1079 Olsen 0.22 0.28 83.7%
Hoffer 2010 Lenstar Acrysof SN60WF 50 Haigis N/A 0.45 58%
Hoffmann 2013 Lenstar 3 IOL models* 308 Holladay 1 0.26 0.31 79.2%
Hoffmann 2013 Lenstar Acrysof SN60WF 82 Holladay 1 0.25 0.31 N/A
Reitblat 2015 Lenstar Restor SN6AD1 73 SRK/T N/A 0.23 91.78%

Accuracy of biometric measurements for intraocular lens power calculation with different optical biometers and formulas (only the results of the best formula are shown).

* Acrysof SN60WF (Alcon), iMics1 (Bausch + Lomb), Tecnis (Abbott Medical Optics).

** the percentage refers only to eyes with an axial length between 22 and 24.5mm.