Table 2.

Linearity, accuracy, and precision data of tenofovir analysis

Linearity data of tenofovir calibration sets
Standard curve Range (µg/mL) Calibrators Slope y-Intercept R2 value
Validation 1 1.0–10 6 0.0566 −0.0691 0.9995
Validation 2 1.0–10 6 0.0579 −0.0948 0.9996
Validation 3 1.0–10 6 0.0603 −0.0166 0.9995

Accuracy and precision: drug substance (n = 6)
Validation parameters Validation solutions 1 µg/mL 2 µg/mL 6 µg/mL 10 µg/mL

Accuracy (%) Set 1 94.6 96.8 101.2 99.4
Set 2 95.5 98.6 100.0 99.5
Set 3 98.0 99.1 100.7 99.8
Precision (%RSD) Set 1 2.58 1.51 0.625 1.15
Set 2 0.84 1.17 0.334 0.49
Set 3 0.98 1.61 0.72 1.70

Abbreviation: RSD, relative standard deviation.