Table 1.

Application on several real ECG signals issued from the standard database MIT-BIH, extraction of QRS complex, sensitivity and specificity rates determination

Record Record length Real number of QRS TP FN FP Se (%) Sp (%) FDR
(%)
FNR
(%)
RR¯ σRR QRS¯ σQRS
100 1805 2273 2272 1 0 99.96 100.00 0.00 0.04 0.79 0.05 0.05 0.00
101 1805 1865 1864 1 0 99.95 100.00 0.00 0.05 0.96 0.07 0.06 0.00
102 1805 2187 2183 4 2 99.82 99.91 0.09 0.18 0.83 0.09 0.14 0.01
103 1805 2084 2082 2 2 99.90 99.90 0.10 0.10 0.86 0.05 0.05 0.00
104 1805 2230 2211 19 24 99.15 98.93 1.07 0.85 0.81 0.08 0.04 0.04
105 1805 2572 2571 1 0 99.96 100.00 0.00 0.04 0.70 0.11 0.07 0.00
106 60 67 67 0 0 100.00 100.00 0.00 0.00 0.88 0.09 0.06 0.00
107 60 70 70 0 2 100.00 97.22 2.78 0.00 0.82 0.12 0.12 0.02
108 1805 1763 1761 2 0 99.89 100.00 0.00 0.11 1.02 0.21 0.09 0.08
109 1805 2532 2526 6 2 99.76 99.92 0.08 0.24 0.71 0.05 0.09 0.00
111 60 69 69 0 0 100.00 100.00 0.00 0.00 0.89 0.15 0.05 0.01
112 1805 2539 2537 2 6 99.92 99.76 0.24 0.08 0.71 0.03 0.06 0.00
113 1805 1794 1794 0 1 100.00 99.94 0.06 0.00 1.00 0.09 0.05 0.00
114 60 54 54 0 0 100.00 100.00 0.00 0.00 1.10 0.04 0.03 0.00
115 1805 1953 1953 0 0 100.00 100.00 0.00 0.00 0.92 0.08 0.05 0.00
116 60 78 78 0 0 100.00 100.00 0.00 0.00 0.76 0.01 0.06 0.00
117 1805 1535 1534 1 1 99.93 99.93 0.07 0.07 1.17 0.05 0.06 0.00
118 1805 2275 2275 0 12 100.00 99.48 0.52 0.00 0.78 0.10 0.07 0.00
119 1805 1987 1987 0 0 100.00 100.00 0.00 0.00 0.90 0.25 0.07 0.02
121 1805 1863 1861 2 3 99.89 99.84 0.16 0.11 0.96 0.09 0.08 0.00
122 1805 2476 2475 1 2 99.96 99.92 0.08 0.04 0.72 0.04 0.07 0.00
123 1805 1518 1517 1 4 99.93 99.74 0.26 0.07 1.18 0.12 0.06 0.00
124 60 49 49 0 0 100.00 100.00 0.00 0.00 1.21 0.02 0.07 0.01
200 60 87 87 0 0 100.00 100.00 0.00 0.00 0.72 0.40 0.09 0.01
201 60 90 90 0 0 100.00 100.00 0.00 0.00 0.66 0.13 0.06 0.00
202 1805 2136 2111 25 0 98.83 100.00 0.00 1.17 0.85 0.30 0.07 0.00
203 60 97 97 0 0 100.00 100.00 0.00 0.00 0.61 0.24 0.08 0.01
205 60 89 89 0 0 100.00 100.00 0.00 0.00 0.66 0.01 0.05 0.00
207 1805 1862 1859 3 0 99.84 100.00 0.00 0.16 0.96 0.21 0.07 0.09
208 60 87 87 0 0 100.00 100.00 0.00 0.00 0.69 0.26 0.07 0.02
209 1805 3004 3002 2 7 99.93 99.77 0.23 0.07 0.59 0.08 0.05 0.00
210 1805 2647 2606 41 9 98.45 99.66 0.34 1.55 0.69 0.13 0.07 0.01
212 1805 2748 2748 0 5 100.00 99.82 0.18 0.00 0.65 0.04 0.06 0.00
213 1805 3251 3243 8 2 99.75 99.94 0.06 0.25 0.55 0.04 0.06 0.01
214 1805 2262 2229 33 0 98.54 100.00 0.00 1.46 0.81 0.22 0.07 0.00
215 1805 3363 3337 26 0 99.23 100.00 0.00 0.77 0.54 0.09 0.06 0.00
217 1805 2208 2206 2 0 99.91 100.00 0.00 0.09 0.88 0.25 0.10 0.01
219 1805 2154 2152 2 0 99.91 100.00 0.00 0.09 0.83 0.22 0.06 0.00
220 1805 2048 2047 1 4 99.95 99.80 0.20 0.05 0.88 0.09 0.05 0.00
221 1805 2427 2400 27 0 98.89 100.00 0.00 1.11 0.75 0.20 0.06 0.01
222 60 75 75 0 0 100.00 100.00 0.00 0.00 0.81 0.12 0.05 0.00
223 60 80 80 0 0 100.00 100.00 0.00 0.00 0.75 0.08 0.07 0.00
228 60 68 68 0 0 100.00 100.00 0.00 0.00 0.86 0.27 0.07 0.01
230 1805 2256 2219 37 0 98.36 100.00 0.00 1.64 0.81 0.18 0.06 0.00
231 60 63 63 0 0 100.00 100.00 0.00 0.00 0.94 0.11 0.06 0.00
232 1805 1780 1747 33 4 98.15 99.77 0.23 1.85 1.03 0.65 0.06 0.00
233 60 94 94 0 0 100.00 100.00 0.00 0.00 0.58 0.13 0.07 0.01
234 1805 2753 2752 1 0 99.96 100.00 0.00 0.04 0.65 0.03 0.06 0.00
Total 58,720 7,3562 73,278 284 92 99.74 99.86 0.14 0.26 0.82 0.13 0.07 0.01

The average sensitivity (Se) rate of the proposed method was 99.74% and the average specificity (Sp) rate was 99.86%. The average False Detection Rate (FDR) rate and the average False Negative Rate (FNR) rate were 0.14 and 0.26% respectively