Skip to main content
. 2016 Oct 12;282(3):892–902. doi: 10.1148/radiol.2016160831

Figure 6a:

Figure 6a:

(a, b) Receiver operating characteristic curves and (c, d) logistic regression graphs show relationship between antenna position and risk of cardiac tissue injury. There is high risk of causing cardiac tissue injury when microwave antenna is positioned less than approximately 4 mm from heart with either antenna orientation (a, b). This risk decreases with increasing antenna distance from heart at both antenna orientations (c, d).