Table II.

Displacement values (mm) of tested anchors. ICONIX anchors did not reach 150 cycles and Q-FIX anchors did not reach 200 cycles, therefore data are not included; standard error (se) at 200 cycles is not described for all anchors as anchors failed before this point

Studied anchor n Mean displacement at 50 cycles, mm (se) Mean displacement at 100 cycles, mm (se) Mean displacement at 150 cycles, mm (se) Mean displacement at 200 cycles, mm Mean total displacement, mm (se) Mean number of cycles at failure (se)
ConMed (New York, New York) Y-Knot RC 5 9.2 (se 1.2) 15.1 (se 2.1) 20.9 (se 3.7) 27.3 23.6 (se 2.2) 151 (se 19)
Smith & Nephew (London, United Kingdom) Q-FIX 5 6.9 (se 2.2) 11.8 (se 3.4) 14.6 (se 4.5) Construct failure prior to data point 20.3 (se 3.3) 161 (se 16)
Stryker (Kalamazoo, Michigan) ICONIX 5 8.4 (se 0.7) 17.9 (se 1.9) Construct failure prior to data point Construct failure prior to data point 22.7 (se 2.2) 112 (se 12)
Zimmer Biomet (Warsaw, Indiana) JuggerKnot 5 10.3 (se 1.5) 22.6 (se 2.5) 29.6 (se 4.8) 22.9 33.7 (se 4.0) 146 (se 19)
Smith & Nephew TWINFIX Ultra PK Suture Anchors 4 7.5 (se 0.5) 12.5 (se 0.3) 17.0 (se 0.7) 21.5 19.7 (se 1.0) 175 (se 9)