. 2017 Feb 17;17(2):394. doi: 10.3390/s17020394

Table 5.

Measuring errors before and after correction of the piezoelectric sensor for step force.

Set Subset Amplitude of Fa/kN Fm (Before Correction) Fc (After Correction)
Amplitude/kN Error Amplitude/kN Error
1 1-1 50 51.06 2.12% 51.97 3.94%
1-2 100 100.38 0.38% 102.15 2.15%
2 2-1 100 104.53 4.53% 103.47 3.47%
2-2 200 211.85 5.93% 205.52 2.76%
2-3 400 416.3 4.08% 379.88 5.03%
3 3-1 200 214.44 7.22% 203.92 1.96%
3-2 400 382.35 4.41% 381.24 4.69%
3-3 500 480.25 3.95% 492.82 1.44%
4 4-1 50 49 2% 48.2 3.61%