. 2017 Feb 27;13:61–69. doi: 10.2147/VHRM.S126285

Table 1.

Validation results for Microlife WatchBP O3® according to the ESH-IP 2010

Validation results – Microlife WatchBP O3
Part 1 ≤5 mmHg ≤10 mmHg ≤15 mmHg Grade 1 Mean mmHg SD mmHg
Pass requirements
 Two of 73 87 96
 All of 65 81 93
Achieved
 SBP 64 86 94 Fail 0.3 7.8
 DBP 62 93 101 Fail −1.9 6.4

Part 2 2/3≤5 mmHg 0/3≤5 mmHg Grade 2 Grade 3

Pass requirements ≥24 ≤3
Achieved
 SBP 22 6 Fail Fail
 DBP 20 6 Fail Fail

Part 3 Result

Fail

Notes: Accuracy is determined by the number differences in these ranges both for individual measurements (Part 1) and for individual subjects (Part 2). To pass, a device must achieve all the minimum pass requirements shown.

Abbreviations: DBP, diastolic blood pressure; ESH, European Society of Hypertension; IP, International Protocol; SBP, systolic blood pressure; SD, standard deviation.