Table 5. Lower limit of detection* of the artus QS-RGQ assay.

Intended concentration (IU/mL) Replicate N Detected N % Detected Probability (%)
20.0 12 12 100.0 100.0
10.0 12 12 100.0 100.0
5.0 12 12 100.0 100.0
2.5 12 11 91.7 91.8
1.0 12 7 58.3 58.5

*The concentration with 95% probability of detection was 2.95 IU/mL according to probit analysis.