Table 4.

Correlations between differences in RNFL thickness measurements (RNFL2-RNFL1) and SS2 and differences in signal strength (SS2-SS1) using Spearman correlations.

Spearman Correlation Coefficient with (RNFL2 - RNFL1) p-value
SS2
All RNFL averages -0.04 0.61
RNFL average < 90 microns 0.15 0.17
RNFL average ≥ 90 microns -0.25 0.03
SS2 - SS1
All RNFL averages 0.20 0.01
RNFL average < 90 microns 0.30 0.004
RNFL average ≥ 90 microns 0.09 0.46
SS2 - adjusted for (SS2-SS1)
All RNFL averages -0.20 0.01
RNFL average < 90 microns -0.09 0.41
RNFL average ≥ 90 microns -0.32 0.006
(SS2-SS1) - adjusted for SS2
All RNFL averages 0.27 0.0004
RNFL average < 90 microns 0.28 0.008
RNFL average ≥ 90 microns 0.22 0.05