Table 1.

Sterilization methods and processing conditions.

Sterilization
modality
General
classification
Temperature
[°C]
Relative humidity
[%]
EO conc.
[mg L−1]
Ethylene
Oxide (100% EO)
Mild 35 35 850
Medium 54 50 712
Severe 54 75 597

Sterilization
modality
General
classification
VHP
[mg L1]
Dwell time
[min]
Total cycle time
[h:m:s]

Vaporized hydrogen
peroxide
Mild 1315.9, 2189.6 7.5 + 7.5 0:47:03
Medium 1995.2, 2939.4 14.0 + 14.0 0:48:59
Severe 2963.2, 7167.2 22.5 + 22.5 1:18:05

Sterilization modality General classification Pre-processing equilibration Delivered dose range [kGy]

Gamma (γ) Mild 20 h @ 21 °C 14.5–16.0
Medium 20 h @ 21 °C 24.0–26.7
Severe 20 h @ 21 °C 50.0–54.6
Electron beam Mild 20 h @ 21 °C 14.0–14.1
Medium 20 h @ 21 °C 24.0–24.7
Severe 20 h @ 21 °C 49.7–52.3