Table 3.

Technical Validation of the NIC Assay

Technical Validation Step Results
Detection range (LLOD-ULOD) 3.1-260 nM
LLOQ 3.2 nM
Intraassay variation 9%
Interassay variation 14%
Dilution recovery in serum 104% (80-115%)
Dilution recovery in plasma EDTA 92% (81-100%)
Spiking recovery in serum 91% (83-105%)
Spiking recovery in plasma EDTA 94% (80-102%)
Freeze-thaw recovery in serum 95% (89-101%)
Freeze-thaw recovery in plasma EDTA 85% (78-101%)
Analyte stability serum 24 h, 4°C/20°C 88% (80-94%)/82% (71-87%)
Analyte stability plasma EDTA 24 h, 4°C/20°C 90% (86-94%)/98% (95-105%)
Interference biotin, low/high 98%/118%
Interference lipemia, low/high 93%/81%
Interference hemoglobin, low/high 105%/101%

Percentages are reported as mean with range shown in brackets.