Table 1.

LC-MS/MS settings and conditions

LC-MS/MS settings and conditions
Mass spectrometer settings
Curtain Gas 15
Ion Spray Voltage −4500v
Dissolvation
Temperature
600°C
GS1 and GS2 20 and 40 respectively
Declustering potential −110v

Collision energy (CE) for different MRM transitions
Succinyl-CoA 866.2> 408 (CE −55ev)
D4-Succinyl-CoA 870.2 > 408 (CE-55ev)
Malonyl-CoA 852.2> 808 (CE −40ev)

Chromatography Gradient conditions

Time
(min)
Flow rate
(ml/min)
Solvent
A%#
Solvent
B%#
0 0.15 70 30
2 0.15 30 70
6.5 0.15 30 70
7.5 0.15 70 30
10 0.15 70 30
#

Solvent A, 10mM ammonium formate in acetonitrile/water (90/10, vol/vol); Solvent B, 10mM ammonium formate in acetonitrile/water (20/80, vol/vol)