Table 9. Overview of datasets that have already been published, that were obtained from instruments that passed an intercomparison exercise at the World Calibration Center for Aerosol Physics (WCCAP) or that were obtained from instruments that were compared to an instrument that passed an intercomparison exercise.
Station |
CCNC |
Mobility particle spectrometer |
Aerosol mass spectrometer |
||||||
---|---|---|---|---|---|---|---|---|---|
Inter-comparison* | publication | available in EBAS† | Intercomparison† | publication | available on EBAS‡ | Inter-comparison | publication | available in EBAS‡ | |
Note that intercomparisons for CCNCs were not introduced before 2016 (denoted as n.a.=not available). |
|||||||||
BRW | n.a. | n.a. | no instrument | ||||||
CES | n.a. | participated, 2009 (ref. 50) | yes | passed, Dec. 201359 | yes | ||||
FIK | n.a. | yes | June 2009 on-site auditingSeptember 2013 passed lab intercomparisonJanuary 2016 passed lab intercomparison | yes | with collocated PM1 filter data | yes | |||
JFJ | n.a. | participated, 2009 (ref. 50) passed, Jul. 2011, May 2014 | yes | does not apply (prototype instrument) | yes | ||||
MEL | n.a. | yes | (ref. 50), 2011 good on-site intercomparison, May/Jun. 2013 passed lab intercomparison, Jul. 2013 installation of new TSMPS after passed intercomparison, Aug. 2013 removal of old TDMPS, agreement between old and new within required boundaries, 2015 passed on-site intercomparison | yes | passed, Dec. 2013 59, and Mar. 2016 | yes | |||
MHD | n.a. | participated, 2009 (ref. 50) recommendations for improvement received Oct. 2012 | yes | does not apply (different instrument type) | 68,§ | yes | |||
NOT | n.a. | n.a. | data not yet processed | ||||||
PUY | n.a. | passed, Oct. 2013 | yes | no instrument | |||||
SEO | n.a. | n.a. | no instrument | ||||||
SMR | n.a. | yes | participated, 2009 (ref. 50) passed, May 2016 | yes | passed, Dec. 2013 (ref. 59), and Mar. 2016 | yes | |||
VAV | n.a. | passed, Oct. 2013 | yes | no instrument |
*the first CCNC intercomparison will take place in fall 2016 at the European Center for Aerosol Calibration.
†reports are available from the European Center for Aerosol Calibration (http://www.actris-ecac.eu/contact.php).
‡as of 24 August 2016 as a primary dataset on http://actris.nilu.no/, archived in EBAS.
§ref. 68 describes the data quality and instrument reliability for the site, but does not discuss the dataset included here as it covers an earlier time period.