Table 4. Comparison of KIM with other real-time prostate tumor localization modalities.

Localization method Kilovoltage intrafraction monitoring Calypso electromagnetic guidance Navotek radioactive fiducial
Accuracy 0.46 mm 0.54 mm (18) 0.89 mm (8)
Imaging dose 61 mSv at 10 Hz None 64 mSv (19)
MRI distortion No Yes No
Information type Image plus 3 points 3 points 1 point
Marker size 1-mm diameter × 3-mm length 1.85-mm diameter × 8-mm length 240-μm diameter × 2-cm length coil that crumples into a smaller size
Rotation information potential Yes Yes No
Intratreatment validation potential Yes (kV/MV triangulation) No No
Image reconstruction potential Yes No No
Accessibility potential Widespread Limited Limited