Table 2.

Intraoperator repeatability measurements.

Parameter/device ICC 95% confidence interval Differencea (mean ± SD) 95% LoA p value
CCT [μm]
 IOLMaster 700 0.999 0.998–0.999 −0.09 ± 2.76 −0.83–0.65 0.808
 Pentacam HR 0.995 0.992–0.997 −0.25 ± 5.18 −1.65–1.14 0.717
 Cirrus HD-OCT 0.995 0.992–0.997 0.91 ± 5.79 −0.65–2.47 0.250
J0 [D]
 IOLMaster 700 0.989 0.983–0.993 0.02 ± 0.11 −0.01–0.04 0.213
 Pentacam HR 0.986 0.978–0.991 −0.02 ± −0.12 −0.05–0.01 0.252
 iDesign 0.994 0.991–0.996 −0.007 ± 0.07 −0.007–0.01 0.468
J45 [D]
 IOLMaster 700 0.974 0.959–0.984 0.009 ± 0.08 −0.01–0.03 0.406
 Pentacam HR 0.984 0.975–0.990 −0.02 ± 0.06 −0.03–0.001 0.066
 iDesign 0.977 0.963–0.986 −0.01 ± 0.08 −0.03–0.01 0.304
Kmean [D]
 IOLMaster 700 0.999 0.999–1.0 −0.002 ± 0.09 −0.03–0.02 0.842
 Pentacam HR 0.999 0.999–1.0 −0.02 ± 0.08 −0.03–0.005 0.145
 iDesign 0.992 0.987–0.995 0.05 ± 0.29 −0.03–0.13 0.211

CCT: central corneal thickness; J0: astigmatism 90/180 degrees; J45: astigmatism 45/135 degrees; Kmean: mean keratometry; ICC: intraclass correlation coefficient; LoA: limits of agreement; abetween the second measurement of each eye; p values were calculated by one-sample t-test.