. 2017 Mar 5;2017:6181405. doi: 10.1155/2017/6181405

Table 3.

CCT measurement differences between the three devices.

Parameter Device comparison
Pentacam HR
versus
IOLMaster 700
Pentacam HR
versus
Cirrus HD-OCT
IOLMaster 700
versus
Cirrus HD-OCT
CCT [μm]
 Differencea
(mean ± SD)
10.99 ± 7.57 11.44 ± 8.91 0.44 ± 5.12
 Range −11.33–24.67 −24.00–26.67 −12.67–12.33
 95% CI 8.94–13.04 9.03–13.85 −0.93–1.83
p value <0.0001 <0.0001 0.519

CCT: central corneal thickness; CI: confidence interval; abetween the mean measurements of each device and of each eye; p values were calculated by one-sample t-test.