Table 4.
Keratometry measurement differences between the three devices.
| Parameter | Device comparison | ||
|---|---|---|---|
| Pentacam HR versus IOLMaster 700 |
Pentacam HR versus iDesign |
IOLMaster 700 versus iDesign |
|
| J0 [D] | |||
| Differencea (mean ± SD) |
0.033 ± 0.09 | 0.026 ± 0.09 | −0.007 ± 0.07 |
| Range | −0.25–0.27 | −0.16–0.31 | −0.18–0.12 |
| 95% CI | 0.009–0.056 | 0.01–0.05 | −0.03–0.01 |
| p value | 0.009 | 0.041 | 0.426 |
| J45 [D] | |||
| Differencea (mean ± SD) |
−0.011 ± 0.10 | −0.015 ± 0.11 | 0.004 ± 0.06 |
| Range | −0.20–0.23 | −0.29–0.25 | −0.14–0.11 |
| 95% CI | −0.01–0.04 | −0.01–0.05 | −0.01–0.02 |
| p value | 0.412 | 0.325 | 0.591 |
| Kmean [D] | |||
| Differencea (mean ± SD) |
0.49 ± 0.11 | −0.27 ± 0.26 | −0.76 ± 0.25 |
| Range | 0.27–0.91 | −1.34–0.26 | −1.95 to −0.24 |
| 95% CI | 0.46–0.52 | −0.34 to −0.20 | −0.83 to −0.69 |
| p value | <0.0001 | <0.0001 | <0.0001 |
J0: astigmatism 90/180 degrees; J45: astigmatism 45/135 degrees; Kmean: mean keratometry; CI: confidence interval; abetween the mean measurements of each device and of each eye; p values were calculated by one-sample t-test.