Table 2.

Accuracy and reliability of the semi-automated segmentation and volumetry of the kidney parenchyma

N Min. 5th percentile Mean 95th percentile Max. SD p value (two-tailed)
Accuracy (agreement with reference volume) Reader 1 96 −11.8 −10.7 0.4 11.1 21.0 6.2 0.54
Reader 2 96 −11.7 −10.0 −0.1 7.7 9.7 4.8 0.84
Reader 3 96 −12.4 −6.0 0.6 9.9 15.1 4.6 0.17
Total 288 −12.4 −8.4 0.3 9.5 21.0 5.2 0.31
Intra-reader reliability Reader 1 48 −16.4 −12.9 −1.2 6.1 7.4 5.5 0.12
Reader 2 48 −12.8 −11.3 −0.6 7.7 9.3 5.4 0.45
Reader 3 48 −15.3 −10.0 0.7 7.1 8.9 4.7 0.29
Total 144 −16.4 −11.3 −0.4 6.5 9.3 5.2 0.40
Inter-reader reliability Reader 1 vs. 2 192 −14.7 −9.6 0.5 9.6 21.9 5.9 0.26
Reader 1 vs. 3 192 −17.5 −11.0 −0.3 12.7 26.8 7.3 0.63
Reader 2 vs. 3 192 −19.3 −11.3 −0.7 8.3 15.8 5.9 0.08
Total 576 −19.3 −10.7 −0.2 9.8 26.8 6.4 0.52