Table 2. Inter-examiner reproducibility of corneal power measurements by Placido disk-based topography and high-resolution rotating Scheimpflug camera in eyes with keratoconus.

Parameters 1P Sw (D) R (D) CoV (%) ICC (95% CI)
Kf
 Placido 0.06 0.29 0.81 0.65 0.968 (0.936-0.984)
 Scheimpflug 0.56 0.08 0.21 0.17 0.998 (0.996-0.999)
Ks
 Placido 0.05 0.35 0.98 0.74 0.986 (0.972-0.993)
 Scheimpflug 0.74 0.14 0.40 0.30 0.998 (0.995-0.999)
Km
 Placido 0.06 0.31 0.87 0.68 0.980 (0.959-0.990)
 Scheimpflug 0.58 0.09 0.26 0.20 0.998 (0.996-0.999)
Ks-Kf
 Placido 0.27 0.19 0.52 5.97 0.990 (0.981-0.995)
 Scheimpflug 0.34 0.14 0.38 4.70 0.993 (0.986-0.997)

Kf: Flat keratometry; Ks: Steep keratometry; Km: Mean keratometry; Ks-Kf: Astigmatism; D: Diopters; Sw: Within-subject standard deviation; R: Coefficient of repeatability; CoV: Coefficient of variation; ICC: Intraclass correlation coefficient. 1Wilcoxon's signed-rank test.