Skip to main content
. 2017 Mar 18;10(3):453–460. doi: 10.18240/ijo.2017.03.20

Table 3. Agreement between Placido disk-based topography and high-resolution rotating Scheimpflug camera in measuring corneal power in eyes with keratoconus.

Instruments Kf Ks Km Ks-Kf
Placido 44.63±1.65 47.76±3.06 46.12±2.23 3.14±1.92
Scheimpflug 44.99±1.81 47.95±3.05 46.41±2.33 2.96±1.63
1P <0.01 0.03 <0.01 0.26
Difference between instruments -0.37±0.47 -0.19±0.60 -0.29±0.40 0.18±0.66
95% LoA -1.28 to +0.55 -1.36 to +0.99 -1.08 to +0.50 -1.11 to +1.48

Kf: Flat keratometry; Ks: Steep keratometry; Km: Mean keratometry; Ks-Kf: Astigmatism; 95% LoA: 95% limits of agreement. 1Wilcoxon's signed-rank test.

x±s