Table 1.
Technique | Step | Description | T (min) |
---|---|---|---|
Fluorophore | Calibration | Purge tubes of air using N2 gas or sodium dithionite | 0.02–0.05 |
Sample preparation | Dissect tissue (e.g. scalpel, scissors or leaf punch) and place in measuring tube | 0.5–1 | |
Measurements | In general, slopes taken from 1 to 2.5-h. 186 samples per run. Note: more than 186 samples can be simultaneously measured but cycle time between O2 recordings will increase to >6-min, reducing resolution | 0.8 | |
Total | 1.3–1.9 | ||
O2-electrode | Calibration | Prepare and assemble electrodes, including application of membrane and electrode solution. Aerate calibration solutions and obtain zero and saturated O2 values after stabilisation of current | 4–9 |
Sample preparation | Dissect tissue and place inside cuvette and adjust plunger being careful not to introduce air pockets | 1–2 | |
Measurements | Slopes taken after stabilisation of signal and before depletion of O2, usually within 10–40 min but dependent on sample | 20–40 | |
Total | 25–51 | ||
IRGA | Calibration | Change consumables (e.g. soda lime, desiccant, CO2 canister) and zero IRGA chambers | 1–2 |
Sample preparation | Select and clip measuring chamber onto leaf | 0.5–1 | |
Measurements | Allow steady-state gas-exchange to be reached | 10–15 | |
Total | 11.5–18 | ||
MIMS | Calibration | Apply membrane and test membrane stability. Purge tube and inject known volumes of O2 and CO2. Record background consumption | 5–10 |
Sample preparation | Dissect tissue and place inside cuvette and air-seal cuvette | 1–3 | |
Measurements | Allow signal to stabilise (usually 5 min) and record slope between 5 and 20 min | 20 | |
Total | 26–33 |
T (min) represents the estimated time it takes to measure a single sample in minutes. For example, if 20 samples can be measured without recalibration and it takes 20-min to calibrate, then the calibration T is 1-min