Skip to main content
. 2016 Mar 17;61(7):2851–2878. doi: 10.1088/0031-9155/61/7/2851

Table 7.

Recommended measurement settings when using the Hyperion-IID scanner to capture the raw DPC sensor data and approximate values for the expected performance.

Scenario Top Vov Trig Val ΔE/E CRT ∼NECR peak
Low activity (<30 MBq) <15 °C 2.5 V 2/3 28ph ∼12.7% 440 ps/550 ps 280 kcps(40 MBq)
High activity (<60 MBq) <15 °C 2.5 V 2/3 37ph ∼12.7% 440 ps/550 ps 300 kcps(55 MBq)
High temperature <25 °C 2.5 V 4 28ph ∼13.0% 1300 ps Not measured
Large object (∅  40 mm) <10 °C 2.5 V 1 28ph ∼12.8% 260 ps 120 kcps(25 MBq)