Table 1.

The mean and standard deviation (SD) of central corneal thickness (CCT) in three measurements and repeatability measurements obtained with Pentacam and Orbscan

CCT (micron) First Second Third RI p-value* ICC
Pentacam
Grade 1 496.6 ± 32.91 497 ± 32.17 497.5 ± 32.67 12.8 0.758 0.979
Grade 2 459.92 ± 39.55 461.69 ± 40.75 462.69 ± 39.28 9.9 0.015 0.993
Grade 3 433.29 ± 29.53 428.94 ± 41.03 431.88 ± 37.79 24.2 0.742 0.943
Orbscan
Grade 1 487.85 ± 43.46 485.81 ± 40.17 481 ± 42.6 23.6 0.178 0.958
Grade 2 443.78 ± 50.77 442.19 ± 45.87 440.78 ± 47.32 26.3 0.480 0.960
Grade 3 413.25 ± 55.74 413.25 ± 53.8 428.25 ± 51.65 59.3 0.286 0.849

RI: Repeatability index.

ICC: Intraclass Correlation Coefficient.

CCT: Central corneal thickness.

*The p-value was calculated by repeated measure ANOVA.