Table 2.

The mean and standard deviation (SD) of corneal thinnest thickness (CCT) in three measurements and repeatability measurements obtained with Pentacam and Orbscan

Thinnest point (micron) First Second Third RI p-value* ICC
Pentacam
Grade 1 480.87 ± 32.6 480.90 ± 32.42 481.67 ± 32.88 9.6 0.569 0.988
Grade 2 440.96 ± 37.05 440.46 ± 37.66 441.46 ± 36.35 8.0 0.468 0.994
Grade 3 409.43 ± 30.97 403.06 ± 39.69 399.06 ± 37.48 35.7 0.029 0.965
Orbscan
Grade 1 482 ± 44.49 481.93 ± 43.35 480.27 ± 45.59 19.5 0.564 0.975
Grade 2 432.23 ± 50.26 432.42 ± 48.56 432.04 ± 48.93 16.6 0.937 0.984
Grade 3 385.47 ± 50.97 384.89 ± 50.3 389.94 ± 48.17 26.8 0.977 0.956

RI: Repeatability index.

ICC: Intraclass Correlation Coefficient.

*The p-value was calculated by repeated measure ANOVA.