Table 3.

Agreement between paired readings of the central corneal thickness (CCT), and thinnest point in microns with Pentacam and Orbscan

Grade Pentacam
Orbscan
r p-value* p-value** 95% limit of agreement CV
Mean ± SD Mean ± SD
CCT (micron) 1 496.83 ± 30.39 485.7 ± 41.91 0.915 <0.001 0.005 −25.5 to 47.7 3.09
2 462.64 ± 39.37 442.25 ± 47.39 0.954 <0.002 0.000 −9.8 to 50.6 3.97
3 440.58 ± 35.02 418.25 ± 50.97 0.941 <0.003 0.022 −20.0 to 64.6 4.96
Thinnest point (micron) 1 481.14 ± 32.5 481.40 ± 44.11 0.948 <0.001 0.934 −33.3 to 32.8 2.43
2 440.96 ± 36.94 432.23 ± 48.99 0.952 <0.002 0.019 −26.2 to 43.7 3.17
3 400.06 ± 36.17 385.50 ± 49.09 0.892 <0.003 0.017 −31.4 to 60.5 4.87

r: correlation coefficient.

*p-value of correlation coefficient.

**p-value of paired t test.

CV: coefficient of variation.

CCT: Central corneal thickness.