| Mass accuracy | |
|---|---|
| Electrospray survey | |
| Attribute | Value |
| Perform lock spray calibration | Yes |
| Lock spray lock mass | 785.8427 Da/e |
| Lock spray scan | 10 |
| Lock mass tolerance | 0.1 Da |
| MSMS | |
| Attribute | Value |
| Perform lock spray calibration | Yes |
| Lock spray lock mass | 785.8427 Da/e |
| Lock spray scan | 10 |
| Lock mass tolerance | 0.1 Da |
| Noise reduction | |
| Electrospray survey | |
| Attribute | Value |
| Background subtraction type | Normal |
| Background threshold | 35% |
| Background polynomial | 5 |
| Perform smoothing | Yes |
| Smoothing type | Savitzky-Golay |
| Smoothing iterations | 2 |
| Smoothing window | 3 channels |
| MSMS | |
| Attribute | Value |
| Background subtraction type | Normal |
| Background threshold | 35% |
| Background polynomial | 5 |
| Perform smoothing | Yes |
| Smoothing type | Savitzky-Golay |
| Smoothing iterations | 2 |
| Smoothing window | 3 channels |
| Peptide filter | |
| De-isotoping and centroiding | |
| Electrospray survey | |
| Attribute | Value |
| Perform de-isotoping | Yes |
| De-isotoping type | Fast |
| Iterations | 30 |
| Automatic thresholds | No |
| Threshold | 3% |
| Minimum peak width | 4 channels |
| Centroid top | 80% |
| TOF resolution | 10,000 |
| NP multiplier | 0.7 |
| MSMS | |
| Attribute | Value |
| Perform de-isotoping | Yes |
| De-isotoping type | Fast |
| Iterations | 30 |
| Automatic thresholds | No |
| Threshold | 3% |
| Minimum peak width | 4 channels |
| Centroid top | 80% |
| TOF resolution | 10,000 |
| NP multiplier | 0.7 |