. 2017 Mar 24;7:44647. doi: 10.1038/srep44647

Table 2. L18 plan for statistical parameter design in performance optimization of the assessment method.

Run Parameter Results
Drug Conductivity [μS/cm] Signal Frequency [kHz] Electrodes Area [cm2] Electrodes Gap [cm] Signal Amplitude [V] Sensitivity [ml−1] Repeatabi-lity [%] Nonlinea-rity [%] Resolution [ml]
1 526 0.1 1.82 1.4 0.020 5.6 2.99 2.34 0.130
2 526 0.1 3.64 3.0 0.100 4.7 0.05 0.78 0.036
3 526 0.1 7.28 4.6 1.000 3.0 0.16 3.74 0.115
4 526 1.0 1.82 1.4 0.100 4.7 1.02 4.26 0.197
5 526 1.0 3.64 3.0 1.000 4.3 0.34 3.16 0.140
6 526 1.0 7.28 4.6 0.020 3.1 0.03 3.69 0.118
7 526 10.0 1.82 3.0 0.020 4.3 0.89 2.54 0.107
8 526 10.0 3.64 4.6 0.100 2.7 0.32 3.80 0.098
9 526 10.0 7.28 1.4 1.000 2.9 0.05 2.82 0.083
10 666 0.1 1.82 4.6 1.000 4.0 0.80 1.37 0.055
11 666 0.1 3.64 1.4 0.020 4.4 0.82 1.50 0.067
12 666 0.1 7.28 3.0 0.100 2.4 0.09 7.62 0.190
13 666 1.0 1.82 3.0 1.000 6.0 3.25 0.82 0.049
14 666 1.0 3.64 4.6 0.020 3.8 0.54 0.47 0.018
15 666 1.0 7.28 1.4 0.100 6.4 0.95 2.26 0.113
16 666 10.0 1.82 4.6 0.100 3.6 0.61 1.00 0.036
17 666 10.0 3.64 1.4 1.000 6.7 2.15 0.94 0.062
18 666 10.0 7.28 3.0 0.020 2.5 0.10 2.79 0.073
        Overall mean   4.2 0.84 2.55 0.09