Table 1.

Validation parameters in applied analytical methods of trace elements determinations.

Analyte Isotope Operating range [µg/L] V R [%] V r [%] LOD [µg/l] LOQ [µg/l] U [%] Recovery NIST 1643-e [%]
Mn 55 0.20–100 10.4 6.3 0.033 0.198 11.1 108.1
Co 59 0.01–100 4.6 4.6 0.002 0.012 10.0 109.6
Cu 65 0.38–100 5.8 3.5 0.064 0.384 12.4 96.1
Zn 66 2.50–100 11.4 7.6 0.181 1.086 14.6 80.3
Pb 208 0.21–100 13.9 11.3 0.036 0.216 18.0 105.7
Fe 0.30–5000 5.1 3.3 0.09 0.30 13.2 CRM NIES number 13 human hair 98.4

Fe3+ determined by IC; LOD: limit of detection; LOQ: limit of quantification. VR: relative standard deviation of reproducibility. Vr: relative standard deviation of repeatability. nd: no data in CRM certificate. U: expanded uncertainty.