Skip to main content
. 2017 Mar 22;47(Suppl 1):111–128. doi: 10.1007/s40279-017-0691-5

Table 4.

Best practice recommendations for measuring sweat [Na+] of athletes in the field using the absorbent patch technique

Local sweat [Na+]
Conditions Test during exercise (as opposed to passive heat stress or pharmacologically-induced local sweating)
Test in conditions (environment, intensity, season, equipment, etc.) representative of training/competition
Conduct multiple tests within athletes to determine sweat [Na+] in various conditions
Methods Check for background electrolytes in collection system (e.g. patches, storage tubes, etc.)
Anatomical location: consider site accessibility and validity compared with whole-body sweat [Na+] (e.g. forearm may be best suited when considering both factors)
Clean skin immediately prior to application: alcohol, deionized/distilled water rinse, and dry with sodium-free gauze/towel
Apply multiple patches per athlete (e.g. right and left forearm) to have a backup (e.g. in case one patch falls off)
Apply patches 20–30 min after the onset of exercise (to establish steady-state sweating prior to sweat collection)
Avoid hidromeiosis: prevent patch saturation by limiting patch time on skin, using patches with high absorbent capacity, and/or changing patches frequently
Check patches for adherence to skin—flag data if patch becomes detached prematurely
Apply multiple patches per session if expecting significant changes in factors that would impact sweating rate (exercise intensity or environment) or if conditions are conducive to whole-body hidromeiosis (e.g. prolonged intense running in humid, still air)
Avoid cross-contamination when working with multiple athletes (e.g. use clean forcipes for each patch)
Storage Refrigerate (e.g. approximately 4 °C) for up to approximately 3–5 days in airtight (e.g. Parafilm-M® sealed) containers
Analysis IC or ICP-MS in the laboratory; ISE in the field
Analysis in the field recommended if sample storage duration and conditions during transportation cannot be well-controlled
Corrections Use regression equations to predict whole-body sweat [Na+] from local sweat [Na+]
Quality control Flag samples that meet the following criteria:
• Sweat sample volume suggestive of saturated patch (volume depends on specific patch type and size)
• Sweat [Na+] <10 mmol/L or >90 mmol/L
• Sweat [K+] <2 mmol/L or >10 mmol/L

See text for discussion and supporting references

IC ion chromatography, ICP-MS inductively coupled plasma mass spectrometry, ISE ion-selective electrode, [K + ] potassium concentration, [Na + ] sodium concentration