Table 3.

Side-by-side comparison of the mouse bioassay and the mass spectrometry method for BoNT/A in human serum.

BoNT Type A (mLD50) MALDI-TOF MS Mouse Bioassay
Replicate 1 Replicate 2 Mouse 1 Mouse 2
72 Positive Positive Expired Expired
36 Positive Positive Expired Expired
18 Positive Positive Expired Expired
1.8 Negative Negative Expired Survived
0.18 Negative Negative Survived Survived
0.018 Negative Negative Survived Survived
Negative Negative Negative Survived Survived