Table 1. Detection of different tumor components in gbPC-CT, gbCT, clinCT and MRI compared with histopathologic findings.

  gbPC-CT/Histo gbCT/Histo clinCT/Histo MRI/Histo
Fibrous strands 15/15 0/15 0/15 4/15
Pseudocapsule 8/9 0/ 9 0/9 6/9
Calcification 9/17 17/17 4/17 *
Microbleeding 10/11 0/11 0/ 11 *
Diffuse hemorrhage 5/5 3/5 2/5 5°/5
Hyalinization 6/8 6/8 2/8 6/8
Necrosis 2/3 0/3 0/3 1/3

Note. – detection in the different modalities was determined according to histopathological findings.

*Small susceptibility artifacts in susceptibility weighted images (SWI) in 17 samples.

°large hypointense areas.

gbPC-CT – grating-based phase-contrast computed tomography; gbCT – grating-based computed tomography; clinCT – clinical computed tomography; MRI – magnetic resonance imaging; Histo – histological findings.