Left ventricular functional parameters in eight healthy subjects. End-diastolic volume (EDV), end-systolic volume (ESV), stroke volume (SV) and ejection fraction (EF) for proposed 3D cardiac CINE framework and reference multi-slice 2D breath-hold reconstruction are given. The mean values of functional parameters are given together with the associated standard deviation. The P values quantifying the comparison of proposed method with reference fully sampled multi-slice breath hold reconstruction (BH) are also given. The differences between the proposed and reference methods in terms of percentage error are given. No significant differences were found between the values computed with the proposed method and those obtained from the breath-hold multi-slice 2D reference.