Published in final edited form as: Analyst. 2016 Mar 21;141(6):2053–2060. doi: 10.1039/c5an02387d

Table 2.

Accuracy and Precision of the method

Theoretical Cortisol Concentrations (ng/ml)
20.000 0.625
Within Day
Mean Measured Concentration (n=5) 20.422 0.628
Standard Deviation 0.504 0.061
Precision (%RSD) 2.5% 9.7%
Accuracy (% difference) 2.1% 0.5%
Between Day
Mean Measured Concentration (n=15) 21.420 0.531
Standard Deviation 2.631 0.099
Precision (%RSD) 12.3% 18.7%
Accuracy (% difference) 7.1% −15.1%
Extraction Recovery (%, n=5) 97 103
Matrix Spike Recovery (%, n=3) 92

Precision is expressed as % RSD: (standard deviation)/(mean measured concentration)x100; Accuracy is expressed as % difference: [(measured-theoretical conc)/theoretical conc]x100.