Table 2.

List of patients delays (in order of frequency) that occurred at least three times over the six‐month period before the release of PCT. Errors are ranked either high, medium or low for priority for automation due to error frequency and complexity of coding

Error Description Error Frequency Priority for Automation
Treatment machine scheduling • Patient scheduled on a machine that is not compatible with their treatment 19 (16%) High
Prescription • Wrong energy, dose or dose per fraction 18 (15%) High
• Not linked to plan
Incorrect plan status • Treatment plan isn't treatment approved 15 (12%) High
Image guidance document incorrect or missing • Document is not filled out 13 (11%) High
• Document contains errors
Reference point • Point tracking incorrect dose 8 (7%) High
• Dose limits are incorrect
Mislabeled field name • Treatment or imaging field names are incorrect 8 (7%) High
Clearance issue • Field does not clear 8 (7%) High
Incorrect or missing shifts • Move sheet for patient positioning not completed 6 (5%) Medium
• Shifts incorrect
Imaging templates • Templates not included 5 (4%) Low
• Incorrect template attached
Setup fields • Missing match anatomy 4 (3%) Medium
• Incorrect angle
Bolus • Tray bolus not attached to field or incorrectly attached 3 (2%) High