. 2016 Jul 9;10(2):117–122. doi: 10.1016/j.optom.2016.04.005

Table 2.

Comparison of the GAT, RT, and DCT derived IOP readings.

Difference P
DCT (mm Hg) 15.42 ± 3.31 +3.70 ± 2.11 <0.001
GAT (mm Hg) 11.72 ± 2.59
DCT (mm Hg) 15.42 ± 3.31 +6.08 ± 2.84 <0.001
RT (mm Hg) 9.34 ± 3.29
GAT (mm Hg) 11.72 ± 2.59 +2.38 ± 1.92 <0.001
RT (mm Hg) 9.34 ± 3.29

GAT, Goldmann applanation tonometer; RT, rebound tonometer; DCT, dynamic contour tonometer; IOP, intraocular pressure.