Table 2. AI and CTDIvol at Different Tube Voltages (n = 24).
Tube Voltage (kVp) | AI (mean ± SD) | P* | CTDIvol (mGy) | |
---|---|---|---|---|
Non-O-MAR | O-MAR | |||
140 | 27.03 ± 11.53 | 13.14 ± 2.97 | < 0.001 | 11.6 |
120 | 35.72 ± 20.61 | 15.97 ± 4.45 | 0.001 | 7.9 |
100 | 41.35 ± 13.09 | 20.26 ± 6.31 | < 0.001 | 4.7 |
80 | 71.71 ± 46.53 | 24.49 ± 7.34 | < 0.001 | 2.3 |
*Comparison of AI between paired O-MAR and non-O-MAR images. AI = artifact index, CTDIvol = volume CT dose index, O-MAR = orthopedic metal artifact reduction