Skip to main content
. 2017 Apr 2;2017:1892725. doi: 10.1155/2017/1892725

Table 2.

Results of the study.

Capsule pMDI Jet-type nebulizer
C-28
Mesh-type nebulizer
NE-U22
MMAD (μm) 50.0 5.2 3.2 5.2
Distribution pattern Bimodal Unimodal Unimodal Unimodal
Percentage of particles 2–6 μm 17.0% 51.8% 49.0% 40.4%
Standard dose per one-time use (mg) 20 2 20 20
Amount deposited in the bronchi by administration of treatment dose (mg) 3.4 1.0 9.8 8.1