Table 4.

Performance Characteristics of the Traditional Upper Reference Limit for the LVAD Outflow Cannula Peak Velocity Compared With the Proposed New Reference Limits

Characteristic Traditional Reference Limit (2 m/s) Proposed Reference Limit (2.7 m/s for HM2; 3.4 m/s for HVAD)
False positives, n 24 0
True positives, n 4 4
False negatives, n 0 0
True negatives, n 26 50
Sensitivity, % 100 100
Specificity, % 52.0 100
Negative predictive value 100 100
Positive predictive value 14.3 100

Abbreviations as in Tables 2 and 3.