Skip to main content
. 2016 Dec 9;31(4):593–600. doi: 10.1038/eye.2016.277

Table 2. Area under the receiver-operating characteristic curvesa of the structural and functional parameters to diagnose glaucoma.

Sector GCIPL thickness SAP sensitivity MP sensitivity P_1 P_2 P_3
Superotemporal 0.76 (0.62–0.86) 0.66 (0.53–0.78) 0.66 (0.53–0.79) 0.21 0.19 0.99
Superior 0.68 (0.54–0.81) 0.62 (0.47–0.73) 0.61 (0.46–0.74) 0.44 0.38 0.93
Superonasal 0.58 (0.43–0.70) 0.59 (0.45–0.72) 0.59 (0.46–0.72) 0.94 0.91 0.98
Inferonasal 0.67 (0.54–0.79) 0.63 (0.50–0.75) 0.65 (0.52–0.75) 0.64 0.79 0.74
Inferior 0.77 (0.66–0.87) 0.71 (0.58–0.82) 0.70 (0.56–0.82) 0.50 0.35 0.80
Inferotemporal 0.79 (0.67–0.89) 0.71 (0.58–0.82) 0.72 (0.58–0.83) 0.27 0.29 0.93
Average 0.74 (0.61–0.85) 0.68 (0.56–0.79) 0.66 (0.53–0.79) 0.40 0.29 0.81

Abbreviations: GCIPL, ganglion cell-inner plexiform layer; MP, microperimetry; P_1, P-value associated with the comparison of the AUCs between GCIPL and SAP parameters; P_2, P-value associated with the comparison of the AUCs between GCIPL and MP parameters; P_3, P-value associated with the comparison of the AUCs between SAP and MP parameters; SAP, standard automated perimetry.

a

95% Confidence intervals within parentheses.