Table 3. Sensitivity at 95% specificitya of the structural and functional parameters to diagnose glaucoma.

Sector GCIPL thickness SAP sensitivity MP sensitivity P_1 P_2 P_3
Superotemporal 59% (41–77) 20% (05–49) 21% (12–39) <0.001 <0.001 0.85
Superior 38% (18–58) 21% (07–41) 11% (03–27) 0.001 <0.001 0.01
Superonasal 31% (16–49) 16% (06–35) 08% (02–19) 0.001 <0.001 0.02
Inferonasal 41% (11–63) 18% (03–40) 30% (18–48) <0.001 0.06 0.04
Inferior 57% (26–76) 34% (13–54) 38% (23–60) <0.001 0.002 0.52
Inferotemporal 43% (16–77) 31% (14–53) 33% (18–50) 0.04 0.10 0.74
Average 51% (33–71) 25% (02–40) 16% (06–30) <0.001 <0.001 0.06

Abbreviations: GCIPL, ganglion cell-inner plexiform layer; SAP, standard automated perimetry; MP, microperimetry; P_1, P-value associated with the comparison of the sensitivities between GCIPL and SAP parameters; P_2, P-value associated with the comparison of the sensitivities between GCIPL and MP parameters; P_3, P-value associated with the comparison of the sensitivities between SAP and MP parameters.

a

95% Confidence intervals within parentheses.