Table 1.

Volume of blood collected using pre-packaged blood-collection devices (inverted cup—10 devices per lot tested)

Test lot # Mean SD Min Max % variation
1 (Uganda) 2.71 0.56 1.75 3.50 −46
1 (Nigeria) 3.28 0.63 2.27 4.20 −34
2 3.98 0.61 3.07 5.11 −20
3 (Uganda) 3.93 0.49 3.33 4.70 −21
3 (Nigeria) 4.76 0.33 4.39 5.31 −5
4 5.33 0.49 4.29 6.02 7
5 5.46 0.53 4.69 6.12 9
6 5.43 0.62 4.80 6.43 9

All manufacturers specify a 5 μl fingerstick blood sample. % variation is based on the difference between this 5 μl standard and the mean volume in column 2. Lots 1 and 3 were sampled from both Uganda and Nigeria; all other lots were sampled from Nigeria only